This is a legacy product and is no longer available for sale. We continue to service and support this product.
identiFINDER
The identiFINDER instrument series are user-friendly instruments that identify man-made and natural radionuclides and combines high sensitivity with a wide dose rate range. The instrument has a dual purpose design to facilitate locating missing or offending sources and then identifying the source via its gamma spectrometry and nuclide identification capability. The identiFINDER is a complete digital gamma spectroscopy and dose rate system. It integrates multi-channel analyzer, amplifier, high voltage power supply, and memory with an integral scintillation and GM detector. The identiFINDER is ideally suited for homeland security, industrial, medical, nuclear power generation and nuclear fuel cycle applications.
Specifications
Functions |
Nuclide identification, spectrum analysis, dose rate calculation, total dose display, source finding |
---|
Integrated Electronics |
Multi-Channel-Analyzer, PMT preamplifier, spectroscopy amplifier, power supply |
---|
Full specifications
Stabilization |
In the non-Ultra models, a built-in 137Cs reference source (<15 nCi/500 Bq) is used for online stabilization and in-situ calibration without user interaction. With either stabilization method this special identiFINDER feature allows operation over temperatures between -20°C and +55°C (-4°F and +122°F). The Ultra provides the user with the benefits of LED stabilization technology with or without a small 3nCi Cs137 source, as preferred for implementation and transportation |
---|
Software |
The identiFINDER provides for easy storage of up to 100 spectra and rapid transfer to a PC for reach-back and/or qualitative in-situ analysis with the software supplied. |
---|
User Selectable Nuclide Library |
There are 74 reference spectra of radionuclides stored in six libraries: Nuclear, Industrial, Medical, U.S. Customs Service, OSI (On Site Inspection by IAEA officials), and Security. All sub-libraries except OSI can be edited by adding or deleting specific nuclides from the list. Ten reference spectra can be measured by the user and added to the predefined library spectra. Identification is done by a template-matching correlation procedure. |
---|
Physical Dimensions
|
---|
Weight |
1250 g (2.75 lb) with 35.5 mm x 50.8 mm (1.4 in x 2 in) Nal and batteries |
---|
Temperature Range |
-20°C to +55°C (-4°F to +122°F) |
---|
Protection |
Water proof, dust tight |
---|
Protection Class |
IP 54 |
---|
Battery Life |
10-12 hours; common batteries can also be used (i.e., emergency situations) |
---|
Dimensions |
248.92 mm x 94 mm x 76.2 mm (9.8 in x 3.7 in x 3 in) |
---|
Spectrometry System Specifications
|
---|
HV-Bias |
200 V to 1275 V automatically set to suit individual detector |
---|
Shaping Type |
Digital filter |
---|
INL, Top 99% |
<0.05% |
---|
DNL, Top 99% |
<0.1% |
---|
Spectrum Length |
1024 channels |
---|
Pileup Rejection |
<100 nS, pulse pair res. |
---|
Throughput Rate |
>100,000 cps |
---|
Input Rate |
>350,000 cps |
---|
Spectrum Memory |
100 spectra at 1024 channels |
---|
Real Time Presets |
1 s to 1,000,000 s |
---|
Live Time Presets |
1 s to 1,000,000 s |
---|
Dose/ Dose Measurement Specifications
|
---|
Sensitivity |
>10,000 cps/mrem for 35.5 mm x 50.8 mm (1.4 in x 2 in) Nal (TI) detector |
---|
Dose-rate Range |
1 µrem/h to 100 rem/h max dose-rate equivalent to max dose-rate of NAl |
---|
Dose Range |
10 µrem to 100 rem |
---|
Energy Range |
Nal: 15 keV to 3.0 MeV; GM: 60 keV to 1.6 MeV |
---|
Alarm Levels |
Gamma: Four preset levels |
---|
Neutron |
Separate neutron alarms for 3He model, with blue indicator lamp |
---|